American Weigh Scales (AWS) CG2-500 User Manual

Page 4

Advertising
background image

press [ PCS ] to confirm.

5. You may now place any number of items on the

platform, and the scale will display a count.

IV. CALIBRATION

Calibration is NOT required before first use. Only

calibrate if the scale is not weighing properly.

**The CG2-500 calibrates with a 500g weight. The

CG2-100 calibrates with a 100g weight (not included)**

1. Turn the scale ON.

2. Hold the [

MODE

] key for a few seconds. The

display will show "CAL", followed by the required

calibration weight.

3. Place the required calibration weight onto the

platform and press [ ].

4. The display will show "PASS" then return to normal

weighin mode. Calibration is complete!

V. TROUBLESHOOTING

The primary reasons for inaccuracy or malfunction

are low batteries, incorrect calibration, overload or

operating on an unstable surface. Please keep this in

mind and maintain and operate your scale properly.

This scale is a precise instrument and must be

handled with the utmost care and caution.

[Display] Elimination of errors

[OUT2] Recalibrate the scale (Section III).

[EEEE] You are overloading the scale. Remove the

excessive load immediately!

[un2T] Use the scale in a more stable position