Technical description, 1 factory calibration and adjustment, 2 over current protection – BINDER MK 115 User Manual

Page 95: 3 definition of usable volume

Advertising
background image

MKF + MKFT (E3.1) 07/2014

page 95/114

20. Technical description

20.1 Factory calibration and adjustment

This unit was calibrated and adjusted in the factory. Calibration and adjustment were performed using

standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER (certi-

fied since December 1996 by TÜV CERT). All test equipment used is subject to the administration of

measurement and test equipment that is also constituent part of the BINDER QM DIN EN ISO

9001systems. They are controlled and calibrated to a DKD-Standard at regular intervals.

20.2 Over current protection

Alternating climate chambers MKF / MKFT are equipped with an internal protection not accessible from

outside. If these fuses have responded, please contact an electronic engineer or BINDER service.

20.3 Definition of usable volume

The usable volume illustrated below is calculated as follows:

b

c

c

b

a

a

C

B

A

A, B, C = internal dimensions (W, H, D)
a, b, c = wall separation
a = 0.1*A
b = 0.1*B
c = 0.1*C
V

USE

= (A - 2 * a) * (B - 2 * b) * (C - 2 * c)




Figure 24: Determination of the useable volume

The technical data refers to the defined usable volume.

Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them

and thus obtain a homogenous distribution of temperature and humidity.

Advertising
This manual is related to the following products: