Error message

Cause

solution

The device is not work-

ing properly.

Take out the battery, wait

for one minute and then

reinsert the battery. If the

message is shown again,

contact your retailer.

The device cannot be

switched on and will

not switch to standby.

Insert a new battery.

The measured tem-

perature is over 42.2 °C

(108.0 °F).

Check whether the protec-

tive cover for the measur-

ing probe is damaged and

take a new temperature

measurement.

The measured tem-

perature is under 34 °C

(93.2 °F).

Check whether the protec-

tive cover for the measur-

ing probe is clean and take

a new temperature meas-

urement.

Battery low; further

measurements are not

possible.

Replace the battery.

Depleted battery: if

the battery symbol is

shown constantly, no

more measurements

are possible.

Replace the battery.

If you cannot fi nd a solution to your problem here, contact our Cus-

tomer Services.

12. Technical data

Temperature range

34.0 – 42.2 °C (93.2 – 108.0 °F)

Measurement accuracy

In the measurement range of 35.0 to

42.2 °C (95.0 to 108.0 °F): +/-0.2 °C (0.4 °F),

in other indicating ranges: +/-0.3 °C (0.5 °F)

Operating temperature

range

10 – 40 °C (50 – 104 °F), humidity ≤ 95%;

700 – 1060 hPa

Storage temperature range

-25 – 55 °C (-13 – 131 °F), humidity ≤ 95%;

700 – 1060 hPa

Memory

9 memory spaces

Display resolution

0.1

Battery

1 x 3V CR2032 battery

Weight

72 g including battery

Dimensions

153 x 42 x 35 mm

Protective cap

6 pcs

Automatic switch-off

60 seconds

Battery operating life

3,000 measurements in succession or one

year with one to two measurements per

day, including standby mode.

Safety classifi cation

Connected medical device of type

BF

Clinical reproducibility

0.08 °C (< 1 year old), 0.10 °C (1 – 5 years

old), 0.07 °C (> 5 years old)

IP classifi cation

IP 22, protection against harmful ingress of

water and dust