Beurer FT 78 User Manual

Page 69

Advertising
background image

69

Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity
The FT78 Ear thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the FT78

Ear thermometer should assure that it is used in such an environment.
IMMuNITY test

IEC 60601 test level

Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floor should be wood, concrete or ceramic tile. If floors are

covered with shythetic material, the relative humidity should

be at least 30%.

Electrical fast transient/burst

IEC 61000-4-4

±2 kV for power

supply lines

±1 kV air for input/out-

put lines

Not applicable

Mains power quality should be that of a typical commercial or

hospital environment.

Surge

IEC 61000-4-5

±1kV differential mode

±2kV common mode

Not applicable

Mains power quality should be that of a typical commercial or

hospital environment.

Voltage dips, short interrup-

tions and voltage variations

on power supply input lines

IEC 61000-4-11

<5% U

T

(>95% dip in

U

T

for 0.5 cycle

40% U

T

(60% dip in

U

T

for 5 cycles

70% U

T

(30% dip in

U

T

for 25 cycles

5% U

T

(>95% dip in

U

T

for 5 sec

Not applicable

Mains power quality should be that of a typical commercial or

hospital environment. If the use of the FT78 Ear thermometer

requires continued operation during power mains interruptions,

it is recommended that the FT78 Ear thermometer be powered

from an uninterruptible power supply or a battery.

Power frequency (50/60 Hz)

magnetic field IEC 61000-4-8 3 A/m

3 A/m

Power frequency magnetic fields should be at levels characte-

ristic of a typical location in a typical commercial or hospital

environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.