Cleaning the ph probe, Hydrating the ph probe, 0 cleaning the bluelab ph probe – Bluelab Guardian Monitor User Manual

Page 14: 0 hydrating the ph probe, English

Advertising
background image

English

14

14.0 Cleaning the Bluelab pH Probe

To ensure accurate readings the pH probe tip needs to be rinsed in water after each use

and cleaned prior to calibration using the following instructions.
The storage cap must always be put back on after cleaning. Always ensure it contains

enough Bluelab pH Probe KCl Storage Solution to cover the probe tip.

Remove storage cap from pH probe.
Hold the top of the storage cap, twist the cap to

loosen then remove.

Rinse pH probe tip under fresh tap water.
Never use RO (Reverse Osmosis), Distilled or De-

ionized water.

Fill a small plastic container with clean tap water.
Add a small amount of Bluelab pH Probe Cleaner

or mild detergent (dishwashing liquid).

Gently stir the probe tip in the mixture.
Ensure that you do not ‘knock’ the soil pH probe

on the side of the container as this may cause

damage to the probe.
Rinse well under fresh running water to remove all

traces of the detergent mixture.

If the probe tip requires removal of heavy

contamination: Gently brush around the

glassware with a few drops of Bluelab pH Probe

Cleaner or mild detergent (dishwashing liquid) and

a soft toothbrush.

Rinse well under fresh running tap water to

remove all traces of the detergent mixture.

Calibrate pH probe after cleaning, see section

10.0 After calibration, store pH probe in the

storage cap, ensuring there is enough KCl Storage

Solution to cover the probe tip.

1

3

2

4

5

6

7

6

5

3

1

4

2

7

15.0 Hydrating the pH probe

Hydrate the pH probe in Bluelab pH Probe KCl Storage Solution when:
• the probe tip has not always been stored in KCl storage solution, to improve the

reading response speed.

• the probe tip has been accidentally allowed to dry out.

Never use RO (Reverse Osmosis), De-ionized or Distilled water. Pure water changes the chemistry in the reference,
causing the probe to die.

Loosen, then remove the storage cap. Place the pH probe

upright in a plastic container.

Clean the pH probe tip. Ensure the probe tip is cleaned

before hydrating. See section 14.0 for instructions.

Add enough Bluelab pH Probe KCl Storage Solution to a

plastic container to submerge the pH probe tip.

Leave to soak for at least 24 hours. After hydration, always

calibrate the pH probe to ensure accuracy, see section 10.0.

Bluelab pH
Probe KCl
Storage
Solution

1

3

2

4

Advertising