Troubleshooting guide, 0 troubleshooting guide, English – Bluelab Guardian Monitor User Manual

Page 15

Advertising
background image

English

15

16.0 Troubleshooting guide

Trouble

Reason

Correction

Nutrient reading

low

Contaminated conductivity/temperature

probe.

Clean conductivity/temperature probe

(see section 13.0) .

Temperature of conductivity/

temperature probe and solution

different.

Wait 5-10 minutes for conductivity/

temperature probe to reach solution

temperature.

Temperature

reading

inaccurate

Temperature of conductivity/

temperature probe different to solution

temperature.

Wait 5-10 minutes for conductivity/

temperature probe to reach solution

temperature.

pH reading

inaccurate

Contaminated pH probe / glassware

not clean.

Clean pH probe (see Section 14.0); then

calibrate (see Section 10.0).

Using factory default calibration.

Calibrate pH probe (see section 10.0).

Calibration old.

Calibrate pH probe (see section 10.0).

Broken glass bulb, tube or connector.

Check pH probe for damage.

pH probe damaged or old.

Replace pH probe.

Display shows

‘Err’ during

calibration

An attempt was made to calibrate with

pH 4.0 or pH 10.0 calibration solution

more than one hour after calibration

with the pH7.0 solution.

Calibrate to pH 7.0 again then do pH 4.0

/ pH 10.0 calibration within one hour.

Old or contaminated solutions used for

calibration.

Use fresh calibration solutions.

Dirty or contaminated pH probe.

Clean pH probe (see section 14.0).

pH probe tip been allowed to dry.

Hydrate pH probe (see section 15.0).

pH probe damaged or old.

Replace pH probe.

No display

Mains not switched on.

Switch mains power on.

Power adaptor not plugged into the

Guardian Monitor.

Plug power adaptor into the base of the

Guardian Monitor labeled ‘Power Input’.

pH displays ‘or’
pH displays ‘ur’

Over range pH.

Under range pH.

Solution > 14.0 pH.

Solution < 0.0 pH.

Check pH probe connection. pH probe

could be faulty. Guardian Monitor could

be wet inside.

temp displays

‘or’
temp displays

‘ur’

Over range temperature.

Under range temperature.

Solution >51 °C / 122 °F.

Solution <0 °C / 32 °F.

Conductivity/temperature probe or

Guardian Monitor faulty.

conductivity

displays ‘or’

Over range conductivity/nutrient.

Over range conductivity >9.9 EC, 99 CF,

1990 ppm. Conductivity/temperature

probe or Guardian Monitor faulty.

A display shows

_ _ _ _

EC and pH cannot be determined as

temperature range is over/under.
(see technical specifi cation range

section 17.0).

Test pH probe in calibration solutions

and conductivity/temperature probe in

a known conductivity standard solution

to eliminate these as a cause of this

problem. Check solution tank/reservoir

for problems.