Troubleshooting – Bronkhorst PiPS User Manual

Page 8

Advertising
background image

English

8/16

Doc nr.: 9.17.055

Troubleshooting

No output signal of instrument

No power supply

Check power supply

PCB failure

Check cable

Return to factory

Valve blocked

Clean valve (qualified personnel only)

Filter or screen blocked

Clean filter or screen

Sensor failure

Return instrument to factory

Maximum output signal

PCB failure

Return instrument to factory

Sensor failure

Return instrument to factory

High setpoint vs. output offset

PCB failure

Return instrument to factory

Valve blocked

Clean valve (qualified personnel only)

Stoppage

Remove cause

Pressure not correct

Check pressure

Signal lower than expected

System stoppage

Remove cause

Flow is gradually decreasing

(Gas) Condensation

Decrease supply pressure or increase

temperature

Valve adjustment has changed

Check valve adjustment

Controller adjustment wrong

Check controller setting

Oscillation

Input pressure wrong

Adjust pressure

Valve damaged

Check valve

Volume / tubing between

controller and pressure

controller too small/ short

Improve situation

Wrong controller settings

Check controller settings

Small flow occurs when valve is

supposed to be closed

Valve is leaking

Check valve

Parameter not available

Parameter is not supported by

the connected instrument.

Setpoint is neglected

Instrument has wrong controller

mode

Set controller mode to ‘digital input’

Alarm- or counterlimit reached

Reset alarm

Settings menu not accessible

Security mode ‘blocked’ or

‘menu’

Change security mode

Before switching on power, check if all connections have been made according to the hook-up diagram. The hook-

up diagrams are available at the download section of our website: http://www.bronkhorst.com

LED indications can be very useful in case of problems with the instrument. See the “MANUAL INTERFACE: micro-

switch and LED’s” section in the manual “Operation instructions digital instruments (document nr. 9.17.023)”. This

manual is available at the download section of our website: http://www.bronkhorst.com

Contact Bronkhorst High-Tech local sales representative or send e-mail describing your problem to:

[email protected]

Advertising