DCA-23 Dechlorination Analyzer

1.1 Features

•

Uni Strut Rail Mounted System, Easy Installation

•

Plumb and Play Design, Ready to Use

•

Not Flow Sensitive, Eliminates Pressure Regulators and Rotameters

•

Direct Chlorine Readout, pH and Temperature compensated

1.2 Specifications

Sensor and Flow Train

TRO Sensor:

Platinum Indicator/Silver-Silver chloride Reference, PTFE junction

0.75” OD x 10.5” length, Internal Signal Conditioning

pH Sensor:

Glass Indicator/Silver-Silver chloride Reference, PTFE junction

0.75” OD x 10.5” length, Internal Signal Conditioning

Measurement Range:

Chlorine:

0.00 to 2.00 ppm

pH:

0 to 14 pH

Operating Temperature:

0° C to 60° C (32° F to 140° F)

Operating Pressure:

0-50 psig at 60° C

Wetted Materials:

PVC, PP, PTFE, Glass, Platinum, 316 SS

Process Connections:

Input: ¾” NPT Slip fitting,

Drain: ¾” NPT Slip fitting

Response Time:

T90 in 2 minutes

C22 Analyzer:

Measurements:

Chlorine:

0.01 to 2.00 ppm

pH:

0 to 14 pH

Temperature:

0° C to 100° C (32° F to 212° F)

pH Compensation:

Automatic

Display:

4” X 1” LCD, 2 x 16 characters

Enclosure:

NEMA 4X, LxWxD: 5.7” x 5.7” x 7

Outputs:

(1) 4-20 mA for Chlorine (set to Sensors Range, 0-2.00 ppm or 0-1.00 ppm)