About the photometric cell, Installing the batteries – Electro-Chemical Devices (ECD) HCA1 User Manual

Page 6

Advertising
background image

calibration, the Ready Snap® 3 Calibration Verification Standard part #490903) is

available for purchase.

Compliance Verification for Free and Total Chlorine Testing

This DPD test system is accepted by most health departments because this test is USEPA

(DIN Standard 38 408 G4, ISO 7393/2) accepted for testing requirements for Free and

Total Chlorine. The Chlorine Plus meter uses a wavelength of 525nm; and the

compliance requirement is that the colorimeter wavelength is between 490 and 530nm.

The HCA1 Strip Micro CL (DPD-1) uses the same reagents and proportions, and the

resulting solution pH is maintained between 6.2 and 6.5 as specified by AWWA

(American Water Works Association) method 4500-Cl G. It should be understood that

the USEPA does not "approve" commercial DPD delivery systems such as reagent

powder pillows, tablets, dispensers, or HCA1 Strip DPD delivery devices. The HCA1 Strip

Micro CL (DPD-1) for Free Chlorine, and the HCA1 Strip Micro CL (DPD-3) or the HCA1

Strip Micro CL (DPD-4) for Total Chlorine meet your reportable testing requirements

because the HCA1 Strip Micro CL delivers the same chemicals in identical proportions

(see table below); therefore, the system is compliant. Likewise, AWWA proportions are

followed as required for Total Chlorine measurements using Potassium Iodide.

Components (Free Chlorine)

AWWA 4500-CI G

HCA1 DPD-1

Anhydrous DPD sulfate

1.5%

1.5%

Anhydrous Na

2

HPO

4

33.4%

33.4%

Anhydrous KH

2

PO

4

64.0%

64.0%

EDTA

1.1%

1.1%

About the Photometric Cell

The built-in CELL is transparent plastic and, when filled to the top, contains 4ml. The

sturdy CELL design will last for over 20,000 readings. Scratches on the CELL will not

interfere or compromise the accuracy of the readings because of its fixed position. For

best accuracy, rinse cell with clean water immediately after a test is completed. Do not

use solvents, such as acetone, to clean the cell. When the CELL becomes stained or

cloudy from repeated testing, or when the meter does not blank when you press the

ZERO/ON button, the cell needs to be cleaned.

Clean as follows: Fill cell with clean water and move the Cell cleaning brush up-and-

down and back-and-forth along the walls of the cell. Afterwards, rinse the cell and the

meter is ready for use again. Cleaning the cell regularly is especially recommended after

you run a test that is using turbidity or precipitation chemistry for analysis.

Installing the Batteries

1. Unscrew the O-ring sealed battery cover counter-clockwise. Use proper sized

pliers if necessary. Do not disturb the sealing O-ring. Batteries are not

included.

6

Advertising
This manual is related to the following products: