(in sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing

and Reading the sample.

7. The cursor will move across the display while the meter prepares to measure

the sample. Record result displayed (this result is automatically stored in O3

Menu).

After testing is completed, rinse cell immediately.

MENU Cd4 Chlorine Dioxide Test Procedure

1. Remove (1) Micro Strip Cd (DPD-1) and (1) Micro Glycine Strip from the

bottles before beginning the test. Place them in a dry convenient place and

recap the test strip bottles.

2. Press the ZERO/ON button to power the meter on; the display will show all

annunciators, then the current MENU selection, followed by the last reading.

3. Press and re-press the MENU button until the display shows the parameter

Cd4.

4. Rinse the CELL at least 3 times with the water sample you will be testing -

rinsing minimizes the potential for cross contamination from a previous test.

Finally, fill cell to capacity (4mL) with the water sample.

5. Dip the Micro Glycine Strip into the CELL and immediately press READ. This

starts the 20 SECOND countdown timer. During this time move the strip in a

gentle back and forth motion (approx. 2 strokes/Sec). Remove and discard the

strip after “1” on the display disappears. The cursor will move across the

display while the meter prepares to measure the sample. Ignore the result.

6. Press the ZERO/ON button. The cursor will move across the display followed

by 0.00 PPM. Sample is ready for testing.

7. Dip the HCA1 Micro Strip Cd (DPD-1) into the CELL and immediately press

READ. This starts the 20 SECOND countdown timer. During this time move the

strip in a gentle back and forth motion (approx. 2 strokes/Sec). Remove and

discard the strip after “1” on the display disappears. When testing outdoors

(in sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing

and Reading the sample.

8. The cursor will move across the display while the meter prepares to measure

the sample. Record result displayed (this result is automatically stored in Cd4

Menu).

After testing is completed, rinse cell immediately.

MENU PA5 Peracetic Acid Test Procedure

1. Remove (1) Micro PA (DPD-4) Strip from the bottle before beginning the test.

Place it in a dry convenient place and recap the test strip bottle.

2. Press the ZERO/ON button to power the meter on; the display will show all

annunciators, then the current MENU selection, followed by the last reading.

3. Press and re-press the MENU button until the display shows the parameter

PA5.