Logix 3400iq digital positioner calibration – Flowserve 3400IQ User Manual

Page 20

Advertising
background image

Logix 3400IQ Digital Positioner FCD LGENIM3403-00 – 11/07

Logix 3400IQ Digital Positioner Calibration


Re-Cal Button:
Calibration of the Logix 3400IQ digital positioner is very straightforward. The Re-Cal button will perform a
stroke calibration procedure, but the DIP switches are no longer used. The parameter CONTROL_FLAGS
must be configured over the fieldbus network prior to using the Re-Cal button. Once the positioner is
properly configured, press and hold the Re-Cal button for five seconds. The Re-Cal button is only
operational when the FVPTB is in OOS mode.

Stroke Calibration: To perform a stroke calibration, complete the following steps:

•

Set the Transducer block to OOS.

•

Ensure the positioner is configured correctly. Refer to previous section.

•

Check the Periodic Update box under the Custom tab, in the Transducer block window.

•

Ensure air supply and valve tubing matches configuration.

•

Select Perform a stroke calibration procedure option under the CALIBRATE parameter in the NI

Configurator.

•

Click the Write Button.


The valve should now go closed, open, then to a mid-stroke position for a few seconds. CALIBRATE will
return to normal operation at the completion of the calibration process.


Figure 13: Calibration Window


20

Advertising