Maintenance

Functional Test

1. Push cleaning device slowly into mid-position to simulate a malfunction.
2. The LED "Malfunction" lights up and the relay contact "Malfunction" opens.

First error code E 03 is indicated and then error code E 08.

3. Push cleaning device back to its initial position.

The equipment must return to normal operation.

Functional Test Relay Alarm 1, Alarm 2 and Malfunction

1. Press briefly to activate the test mode for 10 sec.

The 7-segment display shows

.

2. Press . While the button is held down LED Alarm 1

lights up and relay contact Alarm 1 opens.

3. Press . While the button is held down LED Alarm 2

lights up and relay contact Alarm 2 opens.

4. Press . While the button is held down the LED

Malfunction lights up and the relay contact Malfunction
opens.

Malfunctions

1. The measuring transducer does not work – no function, no display

Fault:

Equipment fuse defective.

Remedy: Replace equipment fuse with a new one. Check power supply.

Fault:

Power failure. Not all three green LEDs on the base plate are
illuminated at the same time.

Remedy: Check power supply or replace measuring transducer.

2. Measured value increases steadily by 1 ppm per day (or slower) –

ingress of foreign matter can be ruled out

Fault:

Growing contamination of glass cylinder.

Remedy: Operate cleaning device more often.