Fault Finding

Fault Finding

Fault

Possible reason

Remedy

No lighting up of LEDs

Mains fuse defective

Replace mains fuse

Lamp “Malfuncion” on

Basic turbidity of fluid too high

Refer to “Basic turbidity too high” in this table

Glass cylinder highly contaminated

Clean or replace glass cylinder

Glass cylinder scratched (solids in condensate)

Replace glass cylinder. Install ORG 12/ORG 22

acc. to mounting suggestion Fig. 5

Glow lamp in light emitter defective

Replace glow lamp

Cable leading to light emitter defective

Check cable

Calibration of zero point

Cable leading to light receiver defective

Measuring of lamp voltage < 12 V only possible

(0 %) not possible

if lamp intact

Glass cylinder highly contaminated

Clean or replace glass cylinder

Glass cylinder scratched (solids in condensate)

Replace glass cylinder. Install ORG 12/ORG 22

acc. to mounting suggestion Fig. 5

Glass cylinder covered with condensation

Check gaskets of terminal boxes and glass cylinder,

O-rings of light emitter and light receiver

Drying cartridge assumed pink colour ?

Replace drying cartridge

Extremely eccentric position of glow lamp filament

Insert glow lamp with central filament position

Basic turbidity of fluid too high

Add indicated value of 100 % calibration to turbidity
standard value

No indication of measured

Cable leading to light receiver defective

Check cable

value

Basic turbidity too high

Presence of gas or steam bubbles (flashing)

Throttle outlet (indication must drop)

Basic turbidity specific of fluid

To measure basic turbidity: Repeat calibration of 0 %
and 100 % with distilled water

Measured value rises

If ingress of foreign matter can be excluded,

Operate cleaning device more frequently

continually 1 ppm/day

growing contamination of glass cylinder

Glow lamp often defective

Heat accumulation in light emitter since fluid

Remove drying cartridge and install vent nipple

temperature > 60 °C and drying cartridge mounted

If faults occur that are not listed above,
please contact our subsidiary or agency
in your country.