General Information

Parts

1

1-1

CAUTION

Section 1

General Information

1.1 Parts

Phantom Base Assembly

Thin Clip-On Attenuator (X.X mm Cu)

Thick Clip-On Attenuator (Y.Y mm Cu)

1.2 Assembly

Pediatric patient simulation – phantom without attenuators

Small adults – phantom with thin clip-on attenuator

Larger adults – phantom with thick clip-on attenuator

Heavy adults – phantom with both clip-on attenuators

1.3 Measurement Setup

The image intensifier’s grid is either in-place or removed in accordance with the clinical protocol being
simulated.


The entrance surface of the image intensifier is
fragile. Additional attention is required if the grid is
removed.


Fasten the phantom to the entrance surface of the image intensifier. (In some cases, the clips on the
phantom will engage slots on the image intensifier assembly.)

Set systems with variable source to image receptor instance (SID) to a standard value (e.g. 100 cm).

Remove extraneous material from the x-ray beam.

The tabletop and /or patient mattress may be in place if desired.