4 procedure – Fluke Biomedical 07-653 User Manual

Page 7

Advertising
background image

General Information

Procedure

1

1-3

1.4 Procedure

• Operate the system using fluoroscopic and acquisition modes under investigation.
• Note the electrical setting on the control panel (e.g. kV, mA, ms).
• Note the visibility of the small holes in each of the five cross targets.
• Note the visibility of the six steps in each of the two white targets.
• Note the visibility of the two lead (black) targets. These targets establish the black level for the

system.

• Image analysis may be visual, video or digital.


Advertising