Fluke Biomedical 18-216 User Manual

Page 8

Advertising
background image

Nuclear Associates 18-216, 18-216-1000
Operators Manual

2-2

For acceptance testing, it is recommended that both slit and high-contrast resolution pattern adjustments
be made and compared. Then, for routine QC, high-contrast resolution pattern images can be obtained
and compared with those images obtained during acceptance testing.




*Excerpts taken from The American College of
Radiology Mammography Quality Control Manual,
1994 Revised Edition, pages 145-146.