§1020.33 Computed Tomography (CT) Equipment

2

2-1

Section 2

Operation

§1020.33 Computed Tomography (CT) Equipment

Reprinted from: “21CFR 1020.33” pages 535-540.

A.

Applicability

1. The provisions of this section, except for paragraphs (b), (c)(1), and (c)(2) are applicable as specified herein to

CT x-ray systems manufactured or remanufactured on or after September 3,1985.

2. The provisions of paragraphs (b), (c)(1), and (c)(2) are applicable to CT x-ray systems manufactured or

remanufactured on or after November 29, 1984.

B.

Definitions

As used in this section, the following definitions apply:

1) Computed Tomography Dose Index (CTDI) Means

The integral of the dose profile along a line perpendicular to the tomographic plane divided by the product
of the nominal tomographic section thickness and the number of tomograms produced in a single scan;
that is:





Where:

z

=

position along a line perpendicular to the tomographic plane.

D(z) =

Dose at position z.

T

=

Nominal tomographic section thickness.

n

=

Number of tomograms produced in a single scan.

This definition assumes that the dose profile is centered around z=0 and that, for a multiple tomogram
system, the scan increment between adjacent scans is nT.

2) Contrast Scale Means

The change in linear attenuation coefficient per CT number relative to water; that is: