Storage, Battery replacement – Fluke Biomedical 451P User Manual

Page 27

Advertising
background image

Ion Chamber Survey Meter

Maintenance

17

Storage

Storage of Fluke Biomedical instruments must comply with Level B storage

requirements as outlined in ANSI N45.2.2 (1972), Section 6.1.2 (.2). The storage

area shall comply with ANSI N45.2.2 (1972), Section 6.2 Storage Area,

paragraphs 6.2.1 through 6.2.5. Housekeeping shall conform to ANSI N45.2.3

(1972).
Level B components must be stored within a fire resistant, tear resistant, weather

tight enclosure, in a well-ventilated building or equivalent.
Storage of Fluke Biomedical instruments must comply with the following

considerations:
1. Inspection and examination of items in storage must be in accordance with

ANSI N45.2.2 (1972), Section 6.4.1.

2. Requirements for proper storage must be documented and written

procedures, or instructions shall be established.

3. In the event of a fire, post-fire evaluation must be in accordance with ANSI

N45.2.2 (1972), Section 6.4.3.

4. Removal of items from storage must be in accordance with ANSI N45.2.2

(1972), Sections 6.5 and 6.6.

Battery Replacement

 Warning

To prevent possible electrical shock, fire, or personal injury:

Batteries contain hazardous chemicals that can cause

burns or explode. If exposure to chemicals occurs, clean
with water and get medical aid.

Do not disassemble the battery.

Repair the Product before use if the battery leaks.

Be sure that the battery polarity is correct to prevent battery

leakage.

Do not short the battery terminals together.

Do not disassemble or crush battery cells and battery

packs.

Do not keep cells or batteries in a container where the

terminals can be shorted.

Do not put battery cells and battery packs near heat or fire.

Do not put in sunlight.

Do not operate the Product with covers removed or the case

open. Hazardous voltage exposure is possible.

This Product contains lithium cells with a potential voltage

of 105 V on the battery assembly. To prevent personal
injury or damage to the Product, be careful with this
assembly during removal and installation.