Ion Chamber Survey Meter

Maintenance

19

Replaceable Parts Information

Fluke maintains a complete inventory of all normal replaceable parts. To place an

order, or to obtain information concerning replaceable parts, contact Fluke. See

the How to Contact Fluke section in this manual.

Table 3. Replaceable Parts List

Part Number

Description

3567752

Keypad Overlay, Red

3567765 Handle,

Yellow

3567776

Handle, Foam Grip, Black

3463380

Check source, CS-137, 10 UCI, Flat disc, 1-inch diameter

614487

Battery, 9 V, Alkaline

Recalibration and Service Information

Fluke recommends a once yearly calibration on your Survey Meter. If your

Survey Meter needs recalibration or repair, we request that you contact Fluke.

See the How to Contact Fluke section in this manual.
More information concerning the operation, application, or service of your Survey

Meter can be obtained from the applications engineer at the numbers listed

above.

Troubleshooting

Caution

For safe operation and maintenance of the Product, use caution

when working on this Product. If the device indicates a

relatively high radiation rate, there is a possibility that the

Product has been contaminated. High rates can quickly cause

the operator’s cumulative exposure to increase with the

potential for injury.

Note

Install fresh batteries prior to performing any calibration on the
Survey Meter.

If the Survey Meter fails self-test, it will remain locked on the firmware revision

number. See the How to Contact Fluke section.