2 file blocks – GBS Elektronik MCA-527 Binary Data Format User Manual

Page 12

Advertising
background image

MCA527

Position

Name

Data type

References

Byte offset 196 AHRC group 3 width

unsigned long

CMD_QUERY_STATE527_EX2, byte offset 12

Byte offset 200 AHRC group 4 width

unsigned long

CMD_QUERY_STATE527_EX2, byte offset 16

Byte offset 204 AHRC group 5 width

unsigned long

CMD_QUERY_STATE527_EX2, byte offset 20

Byte offset 208 AHRC group 6 width

unsigned long

CMD_QUERY_STATE527_EX2, byte offset 24

Byte offset 212 AHRC group 7 width

unsigned long

CMD_QUERY_STATE527_EX2, byte offset 28

Byte offset 216 AHRC group 8 width

unsigned long

CMD_QUERY_STATE527_EX2, byte offset 32

Byte offset 220 AHRC group 9 width

unsigned long

CMD_QUERY_STATE527_EX2, byte offset 36

Byte offset 224 AHRC trigger threshold

unsigned short CMD_QUERY_STATE527_EX2, byte offset 40

Byte offset 226 Data coding method

6

unsigned short

3.2 File blocks

Basis file block

This file block already exists. See above.

Timestamps

The file block size corresponds to the used memory size (basis file block, byte offset 72).

There are three data coding methods. The program automatically chooses the data coding
method with the minimal storage requirements.

Data coding method 0: The values are stored with varying byte length. At this method the
bytes are arranged in big-endian order. The first byte always contains the code for the byte
length of the value.

1 byte:

XXXXXXXX

0 … 191

2 byte:

11XXXXXX XXXXXXXX

192 … 12,479

3 byte:

1111XXXX XXXXXXXX XXXXXXXX

12,480 … 798,911

4 byte:

111111XX XXXXXXXX XXXXXXXX XXXXXXXX

798,912 ... 67,907,775

Each value represents the number of time units which are bygone since the previous event.
The time unit length is defined in the basis file block. Values of 67,907,775 means that no
event has occurred for 67,907,775 time units. Only the next value less than 67,907,775 marks
the next event.

Data coding method 1: Each unsigned char value represents the number of time units which
are bygone since the previous event. The time unit length is defined in the basis file block.
Values of 255 means that no event has occurred for 255 time units. Only the next value less
than 255 marks the next event.

Data coding method 2: Each unsigned short value represents the number of time units which
are bygone since the previous event. The time unit length is defined in the basis file block.
Values of 65,535 means that no event has occurred for 65,535 time units. Only the next value
less than 65,535 marks the next event.

RS232 received data

This file block exists if the used memory size (basis file block, byte offset 72) is greater that
zero

7

, and 'Extension port part A configuration' (basis file block, byte offset 102) is 'RS232

(TX/RX, with data buffering)' (=5) or 'Extension port part C configuration' (basis file block, byte
offset 104) is 'RS232 (RX, with data buffering)' (=5). The file block is already 1024 bytes long.

Remarks

This file block is optional. It is a freely defined file block.

6 If this parameter is missing in the file, data coding method 2 is assumed.
7 If the used memory size is equal to zero, the file contains no measurement data, and consequently also no RS232 received

data.