CMG-DM24S12AMS Operator's guide
1. Open the Configuration setup window for the digital sensor, and
check Enable External Trigger Output to make it send triggers to
Scream!.
2. UPLOAD the new configuration to the digital sensor.
3. Open the Configuration setup window for one of the
DM24S12AMS's internal digitisers, and check Enable External
Trigger Input to make it listen for triggers coming from the
digital sensor via Scream!, and record data when it receives one
(depending on its Output control configuration.)
4. UPLOAD the new configuration to the internal digitiser.
5. Repeat for the other internal digitiser.
March 2004