Sensor adjustment, Dip switch settings – Legrand CS Series 360° PIR User Manual

Page 6

Advertising
background image

SENSOR ADJUSTMENT

The sensor comes factory preset and ready for operation.
If testing of operation is desired:
• Remove the sensor’s cover (twist).
• Refer to the DIP switch settings chart below for switch configurations.
• Make sure that office furniture and fixtures are in place.
1. Restore power to the lighting circuits and turn the lights on.

• There is a one minute warm-up when power is initially restored to the

sensor before the sensor works properly.

2. Set Time Delay to minimum: DIP switches #1 through 6 set to ON

(15 secs.).

3. Set Sensitivity to maximum: DIP switches

#7 and #8 to ON.

4. Replace the sensor’s cover.
5. Move away from sensor and be still.

The lights should go off after 15 seconds.
Note: If lights remain on, move farther away or out of sight of the sensor.
If lights still remain on, set the Sensitivity to medium high by setting DIP
switches #7 to ON and #8 to OFF, or see Unwanted Sensor Activations
under Troubleshooting.

6. Set the appropriate Time Delay (DIP switches #1 through 6).

The factory preset Time Delay is 18 minutes; other suggestions are:

Offices and conference rooms: 10 minutes.
Classrooms: 10 minutes.
Warehouses: 6 minutes.
Hallways: 4 minutes.

DIP SWITCH SETTINGS

LED

DIP Switch Control

for Time Delay and

Sensitivity

DIP Switch #

15 seconds

4 minutes

6 minutes

8 minutes

10 minutes

12 minutes

14 minutes

16 minutes

18 minutes

20 minutes

22 minutes

24 minutes

26 minutes

28 minutes

2 minutes

Minimum

Medium Low

Medium High

Maximum

1 2 3 4

7 8

Sensitivity

Sensitivity

5 6

30 minutes

= ON

= OFF

= Factory Presets

Advertising