Measurement Computing GPIB-488 User Manual
Page 109

Appendix C
IBERR
GPIB-488
C-3
ENOL—No listening device(s)
Cause:
The routine detected no listeners.
Solution:
Verify the following:
•
Make sure that a device has been addressed properly.
•
Check that at least two-thirds of the devices on the GPIB
bus are turned on, including the specified device.
•
Make sure that all connections are secure.
•
Verify the device(s) GPIB address.
•
Make sure that the device was properly configured.
EADR—GPIB interface board not addressed
Cause:
The GPIB interface board has not been addressed.
Solution:
Call
ibcmd
and address the board.
If you are calling
ibgts
, with shadow handshaking enabled, call
ibcac
.
EARG—Invalid argument
Cause:
An invalid parameter has been provided to a routine.
Solution:
Check syntax and range of parameters. See Chapter 3,
, and Chapter 4,
.
ESAC—Board is not the system controller
Cause:
Routine requires that the GPIB interface board is the system
controller.
Solution:
Configure the board to be the system controller.