Measurement Computing Personal Daq rev.6.0 User Manual
Page 110
Advertising

6-6 Calibration
878395
Personal Daq User’s Manual
22.
Disconnect the calibration equipment from the Personal Daq
•
VDC Calibrator
•
Harness and multimeter
•
copper short (remove from Analog In 2L and Personal Daq common low (COM)
The calibration procedure is complete. You may now return your Personal Daq system to its data
acquisition status.