MicroE Mercury II 5000 User Manual
Page 19

Sensor Alignment and Calibration
IM-Mercury_II_5000_Series Rev. 1
Page 18
©2014 MicroE Systems
Mercury II 5000 Series Encoders
Installation Manual and Reference Guide
Step
Action
6. Perform setup by checking the Calibrate GOP, Calibrate Index, Set Left Limit, and Set Right
Limit boxes.
Note: Not all applications require all steps.
7. Ensure that the Calibration Type is correctly selected for your encoder (linear/rotary sector is
selected above). Press the Start button in Calibration Mode to begin calibration/setup.
Results: Sensor Calibration will begin. Follow the steps in the Calibration Status box to
complete setup of the MII5000 encoder. For example, in the status box above, the instructions
are to move the sensor until reaching 100% complete and then the next calibration step will
start.
8. If there is a failure, turn off the sensor and clean the scale. After cleaning, return to this section.
Start Button
Check Boxes