Seaward 250 PrimeTest User Manual

Page 22

Advertising
background image

PrimeTest 250+

Operating Instructions

20

After 2s the display will return to the start up screen.

If the mains voltage is incorrect then the PrimeTest
250+ will remain in voltage mode until the voltage is
corrected or removed.

3.2 Testing a Class I Appliance

•

Visually inspect the appliance and mains cord

for signs of damage.

• If the appliance passes a visual inspection

proceed with the electrical tests.

•

Plug the earth test lead into the Earth Continuity

Test Terminal on the PrimeTest 250+ end panel.

• Plug the appliance into the PrimeTest 250+

front panel test/mains outlet socket.

• Connect the earth test probe to an exposed

metal part on the appliance.

• If the appliance under test has an ON/OFF

switch, make sure it is in the ON position.

•

Press the CLI test key

• The PrimeTest 250+ will now perform a
measurement of the protective earth

continuity, reverse the test current and repeat

the measurement. The highest measured value

that occurred during the two tests is displayed.

•

If the measured value is higher than the factory

set pass/fail threshold, the measured value

is displayed and the unit indicates a fail result

as shown below and the test sequence is halted.