Output configuration – Super Systems PH2 User Manual

Page 11

Advertising
background image

10

SSi Portable Hydrogen Analyzer

Rev A

Kn can only be calculated by entering the flow of Nitrogen (N

2

), Ammonia (NH

3

) and Dissociated

Ammonia (DA). These values are entered at the bottom of the home screen when Kn is

selected.

To change the display, click on Mode and press Edit. This will allow the operator to choose

between Hydrogen, Dissociation, NH

3

, and Kn for analysis and display. Select the appropriate

mode and press Ok to modify this option.

Minimum H

2

for NH

3

display

There may be times when the process is not active. During these times, the measured

Hydrogen value displayed on the screen will remain valid, but the calculated % NH

3

will not be

valid since it is based on the assumption that the known process gases are involved. In these

cases, it may be more desirable to hide the display of NH

3

instead of displaying a value that is

known to be incorrect. This can be done by setting a minimum level of H

2

that will be present

when the process is active. When the Hydrogen drops below this level, the instrument will know

that the process is not active and the NH

3

readings are not valid so it will not display NH

3

.

Output Configuration

There is one 4-20mA output that can be configured for any of the four parameters. The operator

can select the Source (H

2

, DA, NH

3

, Kn), the zero value (the value to be represented by 4mA) and

the span value (the value to be represented by 20mA).