Super Systems PH2 User Manual
Page 13

12
SSi Portable Hydrogen Analyzer
Rev A
The user will then measure the current at the appropriate output terminals and enter that value
into the blue box by tapping on it once to bring up a number pad. Save this value by pressing
OK.
Click on the Calibrate button to begin the calibration.
Span Calibration - Outputs
To perform a span calibration, click on the Span Output option. The circle will be filled in with a
dot for the selected option. When ready to start the calibration, click on the Prep for Cal button.
The user will then measure the current at the appropriate output terminals and enter that value
into the blue box by tapping on it once to bring up a number pad. Save this value by pressing
OK.
Click on the Calibrate button to begin the calibration.