1. INTRODUCTION

The pulse counter CIP II is designed to work with Tecfluid’s volumetric counters (Covol) and
turbines.
The circuit is based on a microprocessor that, in function of the pulses per unit of volume
programmed, totalizes the volume that flows through the meter and shows this value on the
display. The CIP II has two volume totalizers, one that can be reset (partial) and the other that
cannot be reset (total).
This instrument is battery powered and doesn’t require any wiring. This makes its installation
very easy.

2. OPERATION

To adapt the counter to the meter, the pulses per unit factor specified on the meter must be
programmed. To do this, the plastic front cover must be removed by unscrewing the four
screws on the corners. After that, the push buttons will be accessible.
There are three push buttons with the following functions:

2.1 Programming

By pushing the two push buttons PROG at the same time for at least two seconds, the
display will change to the first stage of the programming mode.

2.1.1 Measuring

Units

In this screen the measuring units should be programmed.

2

There are seven options of units: pounds (lb), cubic metres (m

3

), U.S. Gallons (USGAL),

litres (L), kilograms (kg), grams (g), and without specific units selected (all units off). The
selected unit will be flashing on the display (except when no specific units are selected).
Use the push button marked with the arrow pointing upwards to change from one unit to
the next. By pushing the two push buttons PROG at the same time (do not keep the push
buttons pushed) the selected unit will be saved in memory and the next programming
screen for the pulses per unit will appear. If the RESET push button is pressed then the
the next programming screen for the pulses per unit will appear but the selected unit will
not be saved in memory and the previous measuring unit will be retained.