Rs-485 communications, Tuc-2r troubleshooting with a digital voltmeter – WattMaster WCC III part 10 User Manual

Page 17

Advertising
background image

10. RS-485 COMMUNICATIONS

WCC III Technical Guide

10-15

TUC-2R Troubleshooting with a Digital Voltmeter

Figure 10-19: Typical TUC-2R RS-485 TUC Communications DC Voltage Measurements with the TUC COMM

Loop Connected – LT1785 Driver

Figure 10-20: Typical TUC-2R RS-485 TUC Communications DC Voltage Measurements with the TUC COMM

Loop Connected – 75176 Driver

The voltage measurements in Figure 10-19 are approximate voltages. These voltages are taken when the TUC2R controller is “ON” and
the RS-485 TUC communications loop is connected. These two voltages will fl uctuate slightly and also will momentarily “fl ip” to the
other meter reading’s value. The voltage measurements from “T” to “SHLD” should be around 2.70 VDC. The voltage measurements
from “R” to “SHLD” should be around 3.00 VDC. Typical bad voltage measurement values would be anything above 3.8 VDC and
anything below 1.5 VDC.

The voltage measurements in Figure 10-20 are approximate voltages. These voltages are taken when the TUC2R controller is “ON” and
the RS-485 TUC communications loop is connected. These two voltages will fl uctuate slightly and also will momentarily “fl ip” to the
other meter reading’s value. The voltage measurements from “T” to “SHLD” should be around 2.70 VDC. The voltage measurements
from “R” to “SHLD” should be around 3.00 VDC. Typical bad voltage measurement values would be anything above 3.8 VDC and
anything below 1.5 VDC.

Advertising