Grass Valley Maestro Master Control Installation v.2.4.0 User Manual

Page 320

Advertising
background image

320

MAESTRO — Installation and Service Manual

Section 8 — Software Installation

6.

From this point there are two possibilities:

•

A popup will show that the Processor update was successful.

Repeat

Step

above if another Processor is present. Otherwise,

go to

Step

below.

•

An error message may indicate that the “physical JTAG chain is

broken.” If this message appears, the FPGAs/CPLDs on the Pro-
cessor cannot be updated. Stop the installation process and then
contact Technical Support.

Updating FPGAs/CPLDs on the Hardware Control Panels

Follow these steps to update the FPGAs/CPLDs on the Hardware Control
panels

1.

In the board Status section of the screen, select the hardware control
panel to update.

2.

Right-click on the FPGA field for this control panel. The FPGA/CPLD
update menu will then appear. See

Figure 218

Figure 218. Control Panel FPGA/CPLD Update Menu

This menu shows the names of all FPGAs/CPLDs on the control panel
and the version number of the gateware now running (“loaded”) in
each device. Certain FPGA-type components and all of the CPLD-type
components can be updated using this menu. If a newer (“current”)
version of gateware is available for those components, the menu will
indicate the new version number and display a check box.

3.

Check the “Select All” box.

4.

Select

Update

. You will be asked to confirm the update.

Advertising