58
Compensation-value table for ND 281 B (linear measurement): factory default setting
Compensation-value table
Description:
*
Start character ( * );
ND-281 B
MM
Model of display unit; unit of measure (MM or IN);
SCALING
= 14
Spacing = 14 ( range: 6 – 20)
DATUM
=
+ 0.0000
Datum point 0 mm (value input)
CMP. NR. 00
=
+ 0.0000
Compensation value 0 = 0.000 mm (compensation value 0 is always 0)
CMP. NR. 01
=
----------
Compensation value 1 = no value entered
CMP. NR. 02
=
----------
Compensation values 2 – 63: no values entered (axis is not corrected)
CMP. NR. 03
=
----------
Compensation-value table is empty
CMP. NR. 04
=
----------
CMP. NR. 05
=
----------
CMP. NR. 06
=
----------
CMP. NR. 07
=
----------
.
.
.
.
.
.
.
.
.
.
.
.
CMP. NR. 60
=
----------
CMP. NR. 61
=
----------
CMP. NR. 62
=
----------
CMP. NR. 63
=
----------
*
Final character (*);
Output Format of the Compensation-Value Table