Operating parameters for measured value output
Parameter
Function
P50 RS232
Baud rate
P51 RS232
Number of additional blank lines for
measured value output
Display freeze during measured value output
In operating parameter P23, you can specify how the
measured value output signal will affect the display unit.
Display freeze during measured value output
P23
Concurrent display, no display freeze: The
display value is the current measured value
DISPL. ACTL.
Frozen display: Display is stopped
(frozen) and updated by every
measured value output signal
DISPL. HOLD
Frozen/concurrent display: Display is
frozen as long as a measured value output
signal is present
DISPL. STOP
RS-232-C/V.24 Data Interface (X31)
187 + (11 • number of blank lines)
Measured value output via PRINT function
➤
➤
➤
➤
➤ Press the MOD key (note parameter setting P86).
Duration of measured value transfer
t
D
=
[ s]
baud rate