HEIDENHAIN ND 221 B v.2 User Manual
Page 40

40
Remote Operation over the RS-232-C/V.24 Data Interface
Remote Operation over the RS-232-C/V.24 Interface
You can operate the display unit remotely over the RS-232-C/
V.24 data interface. The following commands are available on
the ND 221 B:
Format:
<ESC>TXXXX<CR> Pressed key
<ESC>AXXXX<CR> Output of screen contents
<ESC>FXXXX<CR> Execute function
<ESC>SXXXX<CR> Special function
Sequence of commands
Meaning
<ESC>T0000<CR>
"0" key
<ESC>T0001<CR>
"1" key
<ESC>T0002<CR>
"2" key
<ESC>T0003<CR>
"3" key
<ESC>T0004<CR>
"4" key
<ESC>T0005<CR>
"5" key
<ESC>T0006<CR>
"6" key
<ESC>T0007<CR>
"7" key
<ESC>T0008<CR>
"8" key
<ESC>T0009<CR>
"9" key
<ESC>T0100<CR>
"CL" key
<ESC>T0101<CR>
"-" key
<ESC>T0102<CR>
"." key
<ESC>T0104<CR>
"ENT" key
<ESC>T0105<CR>
"MOD" key
<ESC>T0107<CR>
"1/2" (datum) key
Sequence of commands
Meaning
<ESC>T1000<CR>
"CE+0" keys
<ESC>T1001<CR>
"CE+1" keys
<ESC>T1002<CR>
"CE+2" keys
<ESC>T1003<CR>
"CE+3" keys
<ESC>T1004<CR>
"CE+4" keys
<ESC>T1005<CR>
"CE+5" keys
<ESC>T1006<CR>
"CE+6" keys
<ESC>T1007<CR>
"CE+7" keys
<ESC>T1008<CR>
"CE+8" keys
<ESC>T1009<CR>
"CE+9" keys
<ESC>A0000<CR>
Output of model designation
<ESC>A0100<CR>
Output of 14-segment display
<ESC>A0200<CR>
Output of current value
<ESC>A0301<CR>
Output of error text
<ESC>A0400<CR>
Output of software number
<ESC>A0900<CR>
Output of highlighted fields
<ESC>F0002<CR>
<ESC>S0000<CR>
Counter RESET
<ESC>S0001<CR>
Lock keyboard
<ESC>S0002<CR>
Release keyboard