1

MODEL 5081-T

SECTION 1.0

DESCRIPTION AND SPECIFICATIONS

SECTION 1.0

DESCRIPTION AND SPECIFICATIONS

• CHOICE OF COMMUNICATION PROTOCOL: HART or F

OUNDATION

Fieldbus.

• LARGE, EASY-TO-READ two-line display shows the process measurement and temperature.

• SIMPLE MENU STRUCTURE.

• ROBUST NEMA 4X and NEMA 7B ENCLOSURE.

• INTRINSICALLY SAFE DESIGN allows the transmitter to be used in hazardous environments

(with appropriate safety barriers).

• NON-VOLATILE MEMORY retains program settings and calibration data during power failures.

• MEASURES CONDUCTIVITY, % CONCENTRATION, PPM, OR CUSTOM CURVE VARIABLE.

• AUTOMATIC TC RECOGNITION simplifies start up.

• AUTOMATIC/MANUAL TEMPERATURE COMPENSATION ensures accurate monitoring and control.

• AUTOMATIC COMPENSATION FOR SENSOR CABLE RESISTANCE improves accuracy of high

conductivity/ low resistivity measurements.

• BUILT-IN PERCENT CONCENTRATION CURVES INCLUDE 0-15% NaOH, 0-16% HCl, 0-30%

and 96-99.7% H

2

SO

4

.

1.1 FEATURES AND APPLICATIONS

The Model 5081-T can be used to measure conductivity
in a variety of process liquids. The 5081 is compatible
with most Rosemount Analytical sensors. See the
Specifications section for details.

The transmitter has a rugged, weatherproof, corrosion-
resistant enclosure (NEMA 4X and IP65) of epoxy-painted
aluminum. The enclosure also meets NEMA 7B explo-
sion-proof standards.

The transmitter has a two-line seven-segment display.
The main measurement appears in 0.8-inch (20 mm) high
numerals. The secondary measurement, temperature
(and pH if free chlorine is being measured), appears in
0.3-inch (7 mm) high digits.

Two digital communication protocols are available: HART
(model option -HT) and F

OUNDATION

Fieldbus (model

options -FF and FI). Digital communications allows
access to AMS (Asset Management Solutions). Use AMS
to set up and configure the transmitter, read process vari-
ables, and troubleshoot problems from a personal com-
puter or host anywhere in the plant.

A handheld infrared remote controller or the HART and
F

OUNDATION

Fieldbus Model 375 communicator can also

be used for programming and calibrating the transmitter.
The remote controller works from as far away as six feet.

Housed in a rugged NEMA 4X and NEMA 7 case, the
5081T measures conductivity or resistivity in the harshest
environments. Transmitter can also be configured, using
the "Custom Curve" feature, to measure ppm, %, or a no
unit variable according to a programmable conductivity vs.
variable curve. The transmitter will automatically recog-
nize the type of RTD (Pt100 or Pt1000) being used.
Measurements are automatically corrected for the resist-
ance of the sensor cable to improve accuracy of high con-
ductivity readings. Temperature compensation choices
are linear slope correction or none (display of raw conduc-
tivity.