Emerson Process Management Micro Motion 3500 Transmitter User Manual

Page 9

Advertising
background image

8

1. Identify the components shown in Figure 5.

Figure 5.

Sensor cable to Model 3500

2. Prepare the cable according to the instructions in Micro Motion’s

9-Wire Flowmeter Cable Preparation and Installation Guide.

3. Ensure that the cable has 360° shielding, continuous from the

transmitter to the sensor’s junction box. Two methods can be used:

Metallic conduit

Shielded or armored cable

Refer to Micro Motion’s 9-Wire Flowmeter Cable Preparation and
Installation Guide
for specific instructions.

4. At the sensor:

a.

Clip the cable drain wires.

b.

Connect wiring inside the junction box housing and tighten the
screws to hold the wires in place.

For information on your sensor’s junction box terminals, see the
sensor installation manual or Micro Motion’s 9-Wire Flowmeter
Cable Preparation and Installation Guide
.

9-wire cable from sensor

Model 3500

Keyed sensor wiring connector

(see Figure 3, page 6)

brown
red

green
white

blue
gray

orange

violet

yellow

black (drain wires
from all wire sets)

brown

yellow

violet

green

blue

red

black (drains)
orange
white
gray

c

c

c

c

c

c

c

c

c

c

c

c

c

c

c

c

2

4

6

8

10

12

14

16

18

20

22

24

26

28

30

32

a

a

a

a

a

a

a

a

a

a

a

a

a

a

a

a

Advertising