Operation – Sealey ADB02 User Manual

Page 2

Advertising
background image

this puts the meter in standby mode and remains stable until

the user operates the ADB02 again.

4.2.1

Partial Reset

the partial register can be reset by

pressing the reset key when the meter

is in standby, meaning when the

display screen shows the word

“TOTAL”.

After pressing the reset key, during

reset, the display screen first of all

shows all the lit-up digits and then all

the digits that are not lit up.

At the end of the process, a display

page is shown with the reset partial

and the reset total.

After a few moments, the reset total

is replaced by the non resettable total.

4.2.2

Resetting the Reset Total

the reset total resetting operation can

only be performed after resetting the

partial register. the reset total can in

fact be reset by pressing the reset key

at length while the display screen

shows reset total as on the following

display page:

schematically, the steps to be taken

are:

1. Wait for the display to show normal

standby display page (with total only

displayed).

2. Press the reset key quickly.

3. the meter starts to reset the partial.

4. While the display page showing the

reset total is displayed. Press the

reset key again for at least 1 second.

5. The display screen again shows all

the segments of the display followed

by all the switched-off segments and

finally shows the display page where

the reset Reset total is shown.

4.3

Dispensing with Flow Rate Mode Display

It is possible to dispense fluids, displaying at the same time:

• the dispensed partial.

• The Flow Rate in [Partial Unit / minute] as shown on the

following display page:

Procedure for entering this mode:

• Wait for the Remote Display to go to standby, meaning the

display screen shows total only.

• Quickly press the cAL key.

• start dispensing.

4. OPERATION

3.2

LCD Display

The “LCD” of the METER features two numerical registers

and various indications displayed to the user only when the

applicable function so requires.

Key:

1. Partial register (5 figures with moving comma FROM 0.1 to

99999) indicating the volume dispensed since the reset

button was last pressed.

2. Indication of battery charge.

3. Indication of calibration mode.

4. totals register (6 figures with moving comma FRoM 0.1 to

999999), that can indicate two types of total:

4.1. General total that cannot be reset (totAL).

4.2. Resettable total (Reset totAL).

5. Indication of total multiplication factor (x10 / x100 ).

6. Indication of type of total, (totAL / Reset totAL).

7. Indication of unit of measurement of totals: L=Litres

Gal=Gallons

8. Indication of Flow Rate mode.

9. Indication of unit of measurement of Partial: Qts=Quarts

Pts=Pints

L=Litres

Gal=Gallons

User Buttons

the ADB02 features two buttons (reset and cal) which

individually perform two main functions and, together, other

secondary functions.

the main functions performed are:

• Reset key - resetting the partial register and resettable

total (reset total).

• cal key - entering instrument calibration mode.

Used together, the two keys permit entering configuration

mode, useful for changing the units of measurements and

calibration factor.

4.1

Daily Use

the only operations that need to be done for daily use are

partial and/or resettable total register resetting.

the user should use only the dispensing system of ADB02.

occasionally the meter may need to be configured or

calibrated. to do so, please refer to the relevant sections.

Below are the two typical normal operation displays.

one display page shows the partial and reset total registers.

the other shows the partial and general total. switchover

from resettable total to general total display is automatic and

tied to phases and times that are factory set and cannot be

changed.

NOTE: 6 digits are available for Totals, plus two icons x

10 / x100. The increment sequence is the following:

0.0 → 99999.9 → 999999 → 100000 x 10 → 999999 x 10 →

100000 x 100 → 999999 x 100

4.2

Dispensing in Normal Mode

normal mode is the standard dispensing.

While the count is made, the partial and

resettable total are displayed at the same

time (reset total).

should one of the keys be accidentally

pressed during dispensing, this will have

no effect.

A few seconds after dispensing has

ended, on the lower register, the display

switches from resettable total to general

total: the word reset above the word total

disappears, and the reset total is

replaced by the general total.

ADB02

Issue no: 1 - 21/06/13

Original Language Version

© Jack sealey Limited

Advertising
This manual is related to the following products: