Maintenance, Fig.4 fig.3 – Sealey ADB02 User Manual

Page 5

Advertising
background image

6. MAINTENANCE

5.5.4

Press the RESET Key for 2 Seconds

We now go on to Direct change of the

calibration factor: the word “Direct”

appears together with the currently

Used calibration factor. In the bottom

left part of the display, an arrow appears

(upwards or downwards) defining the direction (increase or

decrease) of change of the displayed value when subsequent

operations 5.5.5 or 5.5.6 are performed.

5.5.5

Press the RESET Key

Arrow direction changes. the operation

can be repeated to alternate the

direction of the arrow.

5.5.6

Press the CAL Key

the indicated value changes in the

direction indicated by the arrow by one

unit for each press of the cAL key.

By continuous holding of the cAL key

the first 5 units will change slowly and

will then change quickly. If the desired value is exceeded,

repeat the operations from point (5.4.5).

5.5.7

Press the RESET Key for 2 Seconds

ADB02 is informed that the calibration

procedure is finished. Before performing

this operation, make sure the indicated

value is that required.

5.5.8

At the end of the calculation, the new

UsER K FActoR is shown for a few

seconds, after which the restart cycle is

repeated to finally achieve standby.

q

ATTENTION! From now on, the indicated factor will become

the calibration factor used by the meter and will continue to

remain such even after a battery change.

5.5.9

the ADB02 stores the new work calibration

factor and is ready to begin dispensing,

using the UsER K FActoR that has just

been calculated.

5.6

Unit Configuration

some models of meter feature a menu with which the user

can select the main measurement unit, Quarts (Qts), Pints

(Pts), Litres (L), Gallons (Gal).

The combination of the unit of measurement of the Partial

register and that of the totals is predefined according to the

following table:

to choose between the 4 available combinations:

5.6.1 Wait for ADB02 to go to Standby.

5.6.2 Press the CAL and RESET keys together. Keep these

pressed until the word “UNIT” appears on the screen together

with the unit of measurement set at that time (in this example

Litres / Litres).

5.6.3 Press the RESET key to select the desired combination of

unit of measurement, amongst those shown below.

5.6.4 Save the new combination by pressing the CAL key at length.

ADB02 will pass through the start cycle and will then be ready

to dispense in the set units.

q

WARNING! the Resettable total and total registers will be

automatically changed to the new unit of measurement. no

ADB02

Issue no: 1 - 21/11/08

Combination

No

Unit of Measure of

the Partial Register

Unit of Measure of

the Totals Register

1

Litres (L)

Litres (L)

2

Gallons (Gal)

Gallons (Gal)

3

Quarts (Qts)

Gallons (Gal)

4

Pints (Pts)

Gallons (Gal)

new calibration is required after changing the Unit of

Measurement.

ADB02 has been designed to require a minimum amount of

maintenance.

6.1

Battery Replacement

ADB02 is supplied with 2 x 1.5 V. alkaline AAA batteries and

features two low-battery alarm levels.

6.1.1 When the battery charge falls below the

first level on the LcD, the fixed battery

symbol appears. In this condition, ADB02

continues to operate correctly, but the

fixed icon warns the user that it is

ADVIsABLE to change the batteries.

6.1.2 If ADB02 operation continues without

changing the batteries, the second

battery alarm level will be reached

which will prevent operation. In this

condition the battery icon starts to flash and is the only one

to remain visible on the LcD.

q

WARNING! Discard the old batteries in accordance with

local authority regulations.

6.1.3 Press RESET to update all the totals.

6.1.4 Loosen and remove the 4 fixing

screws of the bottom cover (fig.3).

6.1.5 Remove the old batteries (fig.4).

6.1.6 Place the new batteries in the same

position as the old ones.

6.1.7 close the cover again, by positioning

the rubber protection as a gasket.

6.1.8 ADB02 will switch on automatically

and normal operation can be

resumed.

the ADB02 will display the same

Reset Total, the same Total and the same Partial indicated

before the batteries were changed.

After changing the batteries, the meter does not need

calibrating again.

6.2

Cleaning

only one operation is necessary to clean the ADB02.

Remove the ADB02 from the system and carefully clean any

residual elements from the turbine.

Note: If the turbine will not rotate freely, contact your sealey dealer.

q

WARNING! DO NOT use compressed air onto the turbine in

order to avoid damage due to excessive rotation.

fig.4

fig.3

Original Language Version

© Jack sealey Limited

Advertising
This manual is related to the following products: