5-4 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 3081FG

a. Remove the end cap of the probe to expose

the terminal block.

b. Refer to Figure 5-2. Disconnect the four

wires (two oxygen signal wires and two
thermocouple wires) from the terminal
block.

c. Disconnect the reference air and the cali-

bration check gas lines.

d. Unscrew the probe from the stack and

remove.

e. Using a replacement probe, refer to para-

graph 2-2c for mechanical installation in-
structions.

f. Refer to paragraph 2-3b for electrical in-

stallation instructions.

g. Refer to paragraph 2-4 for reference air and

calibration check gas installation instruc-
tions.

THERMOCOUPLE

+

(GY)

THERMOCOUPLE

-(RD)

TERMINAL

BLOCK

CONDUIT

CELL

-(WH)

CELL

+

(BK)

26020004

Figure 5-2. Oxygen Probe Terminal Block