6-4 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 3081FG

%

26020024

Figure 6-5. Fault 1, Open Thermocouple

a. Fault 1, Open Thermocouple

The thermocouple connection is open. The
fault displays as shown in Figure 6-5.

1. Refer to Figure 6-4 and check the

thermocouple wiring connections at
terminals 8 and 10. Ensure the wires
are properly connected.

2. Remove power. Disconnect the ther-

mocouple wires (gray and red) from
terminals 10 and 8. Measure the conti-
nuity across the gray and red thermo-
couple leads. The measurement
should read approximately 1-2 ohms.
Larger values indicate the thermocou-
ple is open.

3. If the thermocouple is open, replace

the oxygen probe per paragraph 5-2.

%

26020025

Figure 6-6. Fault 2, Reversed Thermocouple

b. Fault 2, Reversed Thermocouple Active

The thermocouple connections are re-
versed. The fault displays as shown in Fig-
ure 6-6.

1. Allow adequate time for the oxygen

probe to reach operating temperatures.
Probe temperatures below approxi-
mately 500

°C (932°F) may result in this

fault.

2. Refer to Figure 6-4. Check the gray (to

terminal 10) and red (to terminal 8)
wires for the proper placement.

3. Using a multimeter, measure between

terminals 8(-) and 10(+). If the reading
is negative, the thermocouple wiring is
reversed. Rewire as necessary.

4. If the wiring is correct and the probe is

at operating temperature, then the
transmitter electronics are bad. Re-
place the PC board stack assembly per
paragraph 5-1.