Amprobe ACD-10-TRMS-PLUS ACD-10-PLUS Clamp-Multimeters User Manual
Page 14
11
ELECtRICAL SPECIFICAtION Accuracy at 23 °C ± 5 °C & < 75% R.H.
dC Voltage
Range
Accuracy
400.0 mV
±( 0.3% + 4 digits)
4.000, 40.00, 400.0 V
±( 0.5% + 3 digits)
600 V
±( 1.0% + 4 digits)
NMRR :
>50 dB @ 50/60Hz
CMRR :
>120 dB @ DC, 50/60 Hz, Rs=1 kΩ
Input Impedance :
10 MΩ, 30 pF nominal (1000 MΩ for 400.0
mV range)
Transient protection :
6.5 kV (1.2/50 μs surge)
AC Voltage (50Hz ~ 500Hz)
Range
Accuracy
4.000, 40.00, 400.0 V
±( 1.5% + 5 digits)
600 V
±( 2.0% + 5 digits)
CMRR :
>60dB @ DC to 60 Hz, Rs=1 kΩ
Maximum Crest Factor:
< 1.75 : 1 at full scale & < 3.5 : 1 at half scale
limited to fundamental and harmonics,
that fall within the meter specified AC
bandwidth for non-sinusoidal waveforms
Input Impedance :
10 MΩ, 30 pF nominal
Transient protection :
6.5 kV (1.2/50μs surge)
ACD-10 Plus:
Average Sensing
ACD-10 TRMS-PLUS:
True RMS sensing - 5 % to 100 % of range