Amprobe ACD-22SW Digital-Clamp-Meters User Manual

Page 15

Advertising
background image

10

4.000Me

± (1.5% rdg + 5 dgts)

40.00Me

± (3.0% rdg + 5 dgts)

Open circuit volts: -0.45V dc typical, (-1.2Vdc on 400e range)
Overload protection: 600VDC or AC rms

Capacitance (ACD-23SW only) (Auto Ranging)

Ranges

Accuracy

4.000μF

±(3.0% rdg + 15 dgts)

40.00μF, 400.0μF

±(3.0% rdg + 5 dgts)

4.000mF

±(5.0% rdg + 20 dgts)

Minimum input range: >100nF
Overload protection: 600VDC or AC rms

Temperature (ACD-23SW only)

Ranges

Accuracy

0.0°C to 400.0°C

± (1.0% + 1°C)

-35.0°C to 0.0°C

± (2.0% + 3°C)

32.0°F to 400.0°F

± (1.0% + 2°F)

-30.0°F to 32.0°F

± (2.0% + 6°F)

Sensor type: K-type thermocouple
Overload protection: 30V Max

Diode Test (ACD-22SW only)

Test current: 0.8 mA (approximate)
Accuracy: ± (3.0% rdg + 3 dgts)
Open circuit volts: 3.0 Vdc typical

Continuity

Range: 400e
Resolution:
1e
Audible indication:
Less than 25e
Response time:
500ms
Overload protection: 600VDC or AC rms