Amprobe ACD-6-PRO Clamp-On-Multimeters User Manual

Page 13

Advertising
background image

12

ELECTRICAL SPECIFICATIONS
Accuracy is ±(% reading digits + number of digits) or otherwise specified, at 73

°F ± 9°F (23

O

C ±5

O

C) & less than 75% R.H.


True RMS Model ACD-6 TRMS PRO ACV & ACA clamp-on accuracies are specified from
5% to 100% of range or otherwise specified. Maximum Crest Factor are as specified below,
and with frequency spectrums, besides fundamentals, fall within the meter specified AC
bandwidth for non-sinusoidal waveforms.


DC Voltage
RANGE

Accuracy

400.0 mV

1.0% + 3d

4.000V, 40.00V,
400.0V

1.7% + 3d

600V

2.0% + 4d

NMRR : >50dB @ 50/60Hz
CMRR : >120dB @ DC, 50/60Hz, Rs=1k

Input Impedance : 10M

Ω, 30pF nominal

(1000M

Ω for 400.0mV range)




Audible Continuity Tester
Open Circuit Voltage: 0.4VDC typical
Range: 400.0

Ω; Accuracy: 1.5% + 6d

Audible threshold: between 10

Ω and 120Ω.




Ohms
RANGE Accuracy
400.0

1.5% + 6d

4.000k

Ω, 40.00kΩ,

400.0k

1.0% + 4d

4.000M

1.5% + 4d

40.00M

2.5% + 4d

Open Circuit Voltage : 0.4VDC typical

AC Voltage
RANGE Accuracy
50Hz -- 500Hz
400.0mV

1)

4.0% + 5d

50Hz -- 60Hz
4.000V, 40.00V,
400.0V

2.0% + 5d

60Hz -- 500Hz
4.000V, 40.00V,
400.0V

2.5% + 5d

50Hz -- 500Hz
600V

3.0% + 5d

CMRR : >60dB @ DC to 60Hz, Rs=1k

Input Impedance

: 10M

Ω, 30pF nominal

True RMS model ACD-6 TRMS PRO Crest

Factor:

< 2.5 : 1 at full scale & < 5.0 : 1 at half

scale

1)

Selection by RANGE button manually, and

is specified from AC 40mV (AC 60mV
for True RMS model ACD-6 TRMS PRO)
and up



Diode Tester

Open Circuit Voltage

Test Current

(Typical)

< 1.6 VDC

0.4mA

Advertising