Amprobe AD40B AC40B Mini-Clamp-Meters User Manual

Page 7

Advertising
background image

5

SyMBoLS

N

Battery

Refer to the
manual

Double
insulated

Dangerous
Voltage

F

Direct Current

Earth Ground

B

Alternating
Current

R

Audible tone

Complies with
EU directives

Application
around and
removal from
hazardous live
conductors is
permitted

Warnings and Precautions

This instrument is EN61010-1:2001 and EN61010-2-32

certified for Installation Category III. It is recommended for
use with local level power distribution, appliances, portable
equipment, etc, where only smaller transient overvoltages
may occur, and not for primary supply lines, overhead lines
and cable systems.
This instrument must not be used on uninsulated conductors

at a voltage greater than 600V ac/dc.
Use extreme caution when working around bare conductors

and bus bars.
Do not exceed the instrument overload limits per function

(see specifications) nor the limits marked on the instrument.
Exercise extreme caution when: measuring voltage >60 V

DC or 30 V AC RMS // current >10 mA // AC power lines with
inductive loads // AC power lines during electrical storms //
servicing CRT equipment. High voltages can be lethal and
high voltage transients may occur at any time.
Never measure current while test leads are inserted in the

input jacks.
Always inspect your instrument, test leads and accessories for

signs of damage or abnormality before every use. If abnormal
conditions exist (broken or damaged test leads, cracked case,
display not reading, etc.), do not use.
When making voltage measurements, make sure these

ranges function correctly. Take a reading of a known voltage
first.