Electrical specifications – Amprobe 34XR-A Professional-Digital-Multimeter User Manual

Page 16

Advertising
background image

10

Approvals:

>

LISTED
950Z

P

Safety: Conforms to UL1244; EN61010-1:
Cat II - 1000V / Cat III - 600V; Class 2,
Pollution degree II.

EMC: Conforms to EN61326-1, criteria B
This product complies with requirements of
the following European Community
Directives: 89/ 336/ EEC (Electromagnetic
Compatibility) and 73/ 23/ EEC (Low
Voltage) as amended by 93/ 68/ EEC (CE
Marking). However, electrical noise or
intense electromagnetic fields in the vicinity
of the equipment may disturb the
measurement circuit. Measuring
instruments will also respond to unwanted
signals that may be present within the
measurement circuit. Users should exercise
care and take appropriate precautions to
avoid misleading results when making
measurements in the presence of electronic
interference.

Electrical Specifications

(Accuracy at 23 °C ±5 °C, <75 % relative
humidity)

DC VOLTS

Ranges: 400mV, 4V, 40V, 400V, 1000V
Resolution: 100 µV in 400mV range
Accuracy: ±(0.5 % rdg + 1 dgt)
Input impedance:
400mV: >100 M

; 4V: 10 M

; 40V to

1000V: 9.1 M

Overload protection: 1000 V dc /
750 Vac rms

AC VOLTS true rms

(45Hz - 2kHz)

Ranges: 400m, 4V, 40V, 400V, 750V
Resolution: 100 µV
Accuracy:
±(1.2 % rdg +8 dgts) 45 Hz to 100 Hz on
400mV range
±(1.2 % rdg + 8 dgts) 45 Hz to 500 Hz
±(2.0 % rdg +8 dgts) 500 Hz to 2 kHz
±(2.0 % rdg + 8 dgts) 45 Hz to 1 kHz on
750 V range
Crest Factor:

≤ 3

Input impedance: 400mV: >100 M

;

4V: 10 M

; 40V to 1000V: 9.1 M

AC coupled true rms specified from
5 % to 100 % of range
Overload protection: 1000 V dc or
750 V ac rms

DC CURRENT

Ranges: 400µA, 4000µA, 40mA, 300mA,
10A
Resolution: 0.1µA
Accuracy: ±(1.0 % rdg + 1 dgt) on 400µA to
300mA ranges
±(2.0 % rdg + 3 dgts) on 10A range

Burden voltage:

400

µ

A Range:

1 mV/ 1

µ

A

4 mA Range:

500 mV/ 1 mA

40 mA Range:

10 mV/ 1 mA

300 mA:

8 mV/ 1 mA

10A:

40 mV/ 1 A

Input protection: 0.315A/1000V fast blow
ceramic fuse 6.3×32mm on µA/mA input
10A/1000V fast blow ceramic fuse
10×38mm on 10A input
10A input: 10 A for 5 minutes maximum
followed by a 10 minute cooling period

AC CURRENT true rms (45Hz - 1kHz)

Ranges: 400µA, 4000µA, 40mA, 300mA,
10A
Resolution: 0.1 µA
Accuracy: ±(1.5 % rdg + 8 dgts) on 400µA
to 300mA ranges
±(2.5 % rdg + 10 dgts) on 10A range
Crest Factor:

≤ 3

Burden Voltage: See DC Current
Input protection: 0.315A/1000V fast blow
ceramic fuse 6.3×32mm on µA/mA input
10A/1000V fast blow ceramic fuse
10×38mm on 10A input
10A input: 10 A for 4 minutes maximum
followed by a 12 minute cooling period