SPECIFICATION
Electrical

Measuring Ranges :

0 - 200.0m ohms in steps of 100μ ohm
0 - 2000m ohms in steps of 1m ohm
0 - 20.00 ohms in steps of 10m ohm
0 - 200.0 ohms in steps of 100m ohm
0 - 2000 ohms in steps of 1 ohm

Accuracy :

±0.5% of reading ±2 digits over the
Operating temperature range, -15°C to
+55°C, with the supplied test leads.

Test Current :

1mA => 2000 ohms range.
10mA => 200 / 20 ohms ranges.
100mA => 2000m / 200m ohms ranges

Test Current Accuracy : ±0.1%

Protection Fuses :

Supply = 0.5 A, HBC, 5 x 20mm, DIN
Current = 0.5 A, HBC, 5 x 20mm, DIN
Voltage = 0.5 A, HBC, 5 x 20mm, DIN

Safety : Complied with CE regulation

Mechanical

Case Size : 110mm (H) X 250mm (W) X 190mm (D)

Bump Test : IEC68-2-29

Vibration Test : IEC1010, clause 8.3

Drop Test : IEC1010, clause 8.4

Impact Test : IEC1010, Clause 8.2

Weight : 1.542kg