Amprobe IR-750 Infrared-Thermometer User Manual

Page 23

Advertising
background image

19

TROUBLE SHOOTING

Code

Problem

Action

OL

Target temperature is over range

Select target within specifications

-OL

Target temperature is under range

Select target within specifications

Battery indication

Low battery

Check and/or replace battery

Blank display

Possible dead battery

Check and/or replace battery

Laser does not work

1. Low or dead battery
2. Ambient temperature above 40°C (104°F)

1. Replace battery
2. Use in area with lower ambient
temperature

Beeper beeps
continuously

Check if high/low alarm has been set and if
the measurement value is over the limit

Reset high/low alarm setting or
cancel the limit setting

BATTERY REPLACEMENT

To install or change one 9V battery (see below):

1. Push the button and pull the handle downward to open battery cover.

2. Install the battery noting its correct polarity.

3. Re-install the battery cover.

Battery: 9V 6F22 alkaline battery or equivalent

Advertising