9

12. Click to exit the program
- Click the “START” from main screen to download the memory from logger

SpECIFICATIoNS

Description

TR100-A

TR200-A

Temperature range

-40 to 85°C(-40 to 185°f)

Humidity range

n/A

0 to 100 %RH

Accuracy:

Temperature

± 0.6°C / 1°f (-20 to 50°C / -4 to 122°f),

±1.2°C / 1.2°f all others

Relative Humidity

n/A

±3% (25°C, 10-90%),

±5% all others

Sample points

1K / 2K / 4K / 8K / 12k / 16K

Resolution:

Temperature

0.1°C / 0.1°f

Relative Humidity

n/A

0.1%

LeD response

Red: Hi, Lo Alarm; Yellow: Record

LCD Display

13 x 13 mm

Dimension

75 x 55 x 23 mm (2.95 x 2.16 x 0.9 in)

Housing

ip65

ip54

TRoUBLESHooTINg
Cannot connect and the display won’t turn on
Ans: Make sure you have selected the correct baud rate. Select 9600 baud rate for all models.
Meter is programmed as Magnetic but could not start
Ans: Set up Magnet start in the Download Suite software. Use a magnet on the rear side of
the meter for more than 2 seconds, the logger should be activated and the ReC flashes. if not,
change a new and stronger magnet and try it again.