Troubleshooting – Amprobe TMD90A Dual-Input-Digital-Thermometer User Manual

Page 13

Advertising
background image

8

Troubleshooting

Problem Solution

Thermometer does not turn
on.

• Check battery voltage and

replace discharged battery.

• Verify that battery clips grip

the battery post tightly.

Dashes appear in the T1 and
T2 data screens.

• Insert missing

thermocouple.

Dashes appear in T1 and/or T2
data screens with
thermocouples inserted.

• Measure resistance of

thermocouples to ensure
they are not broken
internally.

• Clean the thermocouple

and restart.

Temperature drifts from a
known value in a controlled
environment

• Verify that thermocouple

type matches the displayed
icon.

• Clean and dry the

thermocouple blades and
allow to air dry.

• Confirm temperature with a

known good thermocouple.
Replace the defective
thermocouple if required.