12
Specifications
T/C Type
Range Resolution
Accuracy
K
-200 to 650
°C 0.1
°C
651 to 1370
°C 1.0
°C
(0.1% rdg + 0.7
°C)
-328 to 1000
°F 0.1
°F
1001 to 2498
°F 1.0
°F
(0.1% rdg + 1.4
°F)
J
-200 to 500
°C 0.1
°C
501 to 760
°C 1.0
°C
(0.1% rdg + 0.7
°C)
-328 to 940
°F 0.1
°C
941 to 1400
°F 1.0
°F
(0.1% rdg + 1.4
°F
T
-200 to 390
°C 0.1
°C
(0.1% rdg + 0.7
°C)
-328 to 730
°F 0.1
°F
(0.1% rdg + 1.4
°F)
R/S
0 to 1000
°C 0.1
°C
1001 to 1760
°C
1.0
°C
(0.3% rdg + 0.7
°C)
32 to 1000
°F 0.1
°F
1001 to 3200
°F 1.0
°F
(0.1% rdg + 1.4
°F)
E
-200 to 380
°C 0.1
°C
381 to 736
°C 1.0
°C
(0.3% rdg + 0.7
°C)
-328 to 720
°F 0.1
°F
721 to 1832
°F 1.0°F
(0.1% rdg + 1.4
°F)
Additional Specifications
OPERATING CONDITIONS
0 to 50
°C (32 to 122 °F) at 0 to 85 % relative humidity
(non-condensing)
Ambient Coefficient
0 to 18
°C and 28 to 50 °C (Ambient temperatures) For each °C
ambient below 18
°C or above 28 °C, add the following tolerance into
the accuracy spec: 0.01% of reading +0.03
°C (0.01% of reading
+0.06
°F).
TMD90A_all.pdf 16